Directory: 01 Fronting Peak
file name: 01_reconstitution_solvent
Note: There was no problems with mobile phase or needle wash or any method changes.
Symptom #
This is from marijuana metabolite. The normal presentation is on the right, the fronting peaks on the left. The peaks are all the same for three MRM.
We don't use a guard column.
Investigation, Troubleshooting Tool and Root Cause #
Investigation and Tools used:
- Q: Check the Maintenance Log.
A: There were 850 injections and we normally change it at 2500. So it doesn’t really look like column degradation. - Q: Is sample overloaded?
A: The right volume was injected. - Q: Check the system suitability tests results.
A: They like normal. - It makes us suspicious of the sample itself.
Root Cause: Samples were reconstituted with 100% acetonitrile.
Reference:
MSACL: LC-MSMS Troubleshooting 101: Tips and Tricks for Getting Started : PART 1
MSACL: The Basics of LC-MSMS Troubleshooting: Tools, Strategy, Cases